Class II Biological Safety Cabinet
- Product Description
Class II Biological Safety Cabinet
medical/laboratory safety cabinet/class ii biological safety cabinet
medical/laboratory safety cabinet/class ii biological safety cabinet is necessary in animalcule lab, especially in the condition
A Biological Safety Cabinet (BSC), also known as a Biosafety Cabinet is mainly used for handling pathogenic biological samples or for applications that require a sterile work zone. A biological safety cabinet creates inflow and downflow of air that provides operator protection.
A biological safety cabinet (BSC) is a primary engineering control used to protect personnel against biohazardous or infectious agents and to help maintain quality control of the material being worked with as it filters both the inflow and exhaust air. It is sometimes referred to as a laminar flow or tissue culture hood.needing protection measure, such as medicine, pharmacy, scientific research and so on.Biosafety Cabinets (BSC), also known as Biological Safety Cabinets, offer personnel, product, and environmental protection through laminar airflow and HEPA filtration for the biomedical/microbiological lab.Class II biological safety cabinet/biological safety cabinet manufactory's main characters: 1. Air curtain isolation design prevents internal and external cross-contamination, 30% of the air flow is discharged outside and 70% of the internal circulation, negative pressure vertical laminar flow, no need to install pipes.
2. The glass door can be moved up and down, can be positioned arbitrarily, is easy to operate, and can be completely closed for sterilization, and the positioning height limit alarm prompts.3. The power output socket in the work area is equipped with a waterproof socket and a sewage interface to provide great convenience for the operator4. A special filter is installed at the exhaust air to control emission pollution.5. The working environment is made of high-quality 304 stainless steel, which is smooth, seamless, and has no dead ends. It can be easily and thoroughly disinfected and can prevent the erosion of corrosive agents and disinfectants.6. It adopts LED LCD panel control and built-in UV lamp protection device, which can only be opened when the safety door is closed.7. With DOP detection port, built-in differential pressure gauge.8, 10° tilt angle, in line with the human body design concept
Model
|
BSC-1000IIA2
|
BSC-1300IIA2
|
BSC-1600IIA2
|
Airflow system
|
70% air recirculation, 30% air exhaust
|
||
Cleanliness grade
|
Class 100@≥0.5μm (U.S. Federal 209E)
|
||
Number of colonies
|
≤0.5pcs/dish·hour (Φ90mm culture plate)
|
||
Inside the door
|
0.38±0.025m/s
|
||
Middle
|
0.26±0.025m/s
|
||
Inside
|
0.27±0.025m/s
|
||
Front suction air speed
|
0.55m±0.025m/s (30% air exhaust)
|
||
Noise
|
≤65dB(A)
|
||
Vibration half peak
|
≤3μm
|
||
Power supply
|
AC single phase 220V/50Hz
|
||
Maximum power consumption
|
500W
|
600W
|
700W
|
Weight
|
210KG
|
250KG
|
270KG
|
Internal Size (mm) W×D×H
|
1040×650×620
|
1340×650×620
|
1640×650×620
|
External Size (mm) W×D×H
|
1200×800×2100
|
1500×800×2100
|
1800×800×2100
|